Set the Amplitude (RMS Value) of Dip/Swell

Repeatedly push  until the Dip/Swell value is shown in edit mode (blue
background).
The value can be set with the numeric keypad and confirmed by pushing the
V (mA, A) softkeys. The value can also be set with the cursor buttons or
Rotary Knob.
Output terminals are disconnected if a new amplitude is set.
Push the Channel softkey to edit the parameters of each output.

Set Dip/Swell Time Periods

Repeatedly push  until the required time period (t1, t2, t3, t4 or t5) is
shown in edit mode (blue background).
The value can be set with the numeric keypad and confirmed by pushing the
ms (s) softkey. The value can also be set with the cursor buttons or Rotary
Knob.
Output terminals are disconnected if a new time is set.
Push the Channel softkey to edit the parameters of each output.
Generate Electric Energy (Available with 6003A/E Energy
Option installed)
The Product can generate exact values of electric energy. The energy function
modes provide output voltage at VOLTAGE OUTPUT HI – LO terminals and
output current at CURRENT OUTPUT HI – LO terminals.
Power setting range:
Voltage setting range:
Current setting range:
Power factor setting range:
Time setting range:
Frequency setting range:

Energy Pulse Output

Energy pulse output is located on the rear panel (BNC connector is in the right
bottom corner). The frequency of pulses is proportional to the energy generated
by the Product and is defined by the energy meter constant ( this can be
changed in the main setup menu).
There are two types of outputs selectable in the setup menu:
open collector (up to 30 V / 100 mA)
open collector with a pull-up resistor 150 Ω connected to +5 V
Frequency range of energy pulse output:
4-28
0.008 VA to 18 kVA (each channel)
1 V ac to 600 V ac, 1 V dc to 280 V dc
8 mA to 30 A (each channel), 90 mA to 90 A
(High I mode)
-1 to +1 (phase 0 ° to +359.99 °)
1 s to 10000 s
DC, 15 Hz to 1 kHz
0.02 Hz to 1 MHz