Emc Information - Eizo Radiforce GX550-P Instructions For Use Manual

Monochrome lcd monitor
Table of Contents

Advertisement

EMC Information

Essential performance of RadiForce series is to display images and operate functions normally.
CAUTION
The RadiForce series requires special precautions regarding EMC and need to be installed, put into
service and used according to the following information.
Do not use any cables other than the cables that provided or specified by us.
Using other cables may cause the increase of emission or decrease of immunity.
Cable Length: Max. 3m
Do not put any portable and mobile RF communications equipment close to the RadiForce series.
Doing so may affect the RadiForce series.
The RadiForce series should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the equipment or system should be observed to verify normal
operation in the configuration in which it will be used.
Anyone who connects additional equipment to the signal input part or signal output parts,
configuring a medical system, is responsible that the system complies with the requirements of
IEC/EN60601-1-2.
Guidance and manufacturer's declaration - electromagnetic emissions
The RadiForce series is intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Emission test
Compliance
RF emissions
Group 1
CISPR11/EN55011
RF emissions
Class B
CISPR11/EN55011
Harmonic emissions
Class D
IEC/EN61000-3-2
Voltage fluctuations /
Complies
flicker emissions
IEC/EN61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The RadiForce series is intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Immunity test
IEC/EN60601 test
Electrostatic
±6kV contact
discharge (ESD)
±8kV air
IEC/EN61000-4-2
Electrical fast
±2kV
transient / burst
for power supply lines
IEC/EN61000-4-4
±1kV
for input/output lines
Surge
±1kV line(s) to line(s)
IEC/EN61000-4-5
±2kV line(s) to earth
Voltage dips, short
<5% U
interruptions and
U
) for 0.5 cycle
T
voltage variations on
40% U
power supply input
for 5 cycles
lines
70% U
IEC/EN61000-4-11
for 25 cycles
<5% U
U
) for 5sec
T
Power frequency
3A/m
(50/60Hz)
magnetic field
IEC/EN61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
28
Appendix
Electromagnetic environment - guidance
The RadiForce series uses RF energy only for its internal function.
Therefore, its RF emission are very low and are not likely to cause any
interference in nearby electronic equipment.
The RadiForce series is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Compliance level Electromagnetic environment - guidance
level
±6kV contact
±8kV air
±2kV
for power supply lines
±1kV
for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
(>95% dip in
<5% U
(>95% dip in
T
T
U
) for 0.5 cycle
T
(60% dip in U
)
40% U
(60% dip in
T
T
T
U
) for 5 cycles
T
(30% dip in U
)
70% U
(30% dip in
T
T
T
U
) for 25 cycles
T
(>95% dip in
<5% U
(>95% dip in
T
T
U
) for 5sec
T
3A/m
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the RadiForce series requires continued
operation during power mains interruptions, it
is recommended that the RadiForce series be
powered from an uninterruptible power supply or
a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Radiforce gx550Radiforce gx550-arRadiforce gx550-ar-p

Table of Contents