Record the respective aim (average) values as the 'Operating levels' on the
9
Drystar 4500 Charts 2a and 2b ('daily Drystar 4500 control chart'). Refer to
'Quality Control Charts'
The calculated 'Operating levels' should be as follows:
Operating Level
Low density
Mid density
High density
These charts will be used for the daily quality test. For more information, refer
10
to
'Performing the daily QC test'
2801E EN 20050215
on page 77.
Value (Macbeth units)
(according IEC 1223-2-4 or better)
0.4 ± 0.05
1.2 ± 0.15
2.0 ± 0.20
on page 57.
D
4500 U
RYSTAR
Advanced operation (key-operator mode)
SER MANUAL
53