AGFA Drystar 4500 User Manual page 10

Hide thumbs Also See for Drystar 4500:
Table of Contents

Advertisement

D
4500 U
RYSTAR
SER MANUAL
10
Introducing the Drystar 4500
In order to reduce the risk of electric shock, do not remove any
covers.
Type B equipment:
Indicates that the Drystar 4500 complies with the limits for type B
equipment.
Supplementary protective earth connector:
Provides a connection between the Drystar 4500 and the potential
equalization busbar of the electrical system as found in medical
environments. This plug should never be unplugged before the
power is turned off and the power plug has been removed.
Intergrounding connector:
Provides a connection between the printer and other equipment
which might exhibit minor ground potential differences. These
differences may degrade the quality of communication between
different equipment. Never remove connections to this terminal.
Protective earth (ground):
Provides a connection between the printer and the protective earth
of the mains. Do not remove this connection, because this will have a
negative influence on the leakage current.
Power/Reset Button (Standby switch)
Note that the power cord has to be disconnected from the wall outlet
in order to disconnect the unit entirely from the mains.
Precautions for use in USA only:
Make sure that the circuit is single-phase center-tapped, if the printer
is connected to a 240 V/60 Hz source instead of a 120 V/60 Hz
source.
2801E EN 20050215