2.3. Configuration mode

The configuration mode alows you to modify the pump configuration manually. Please
refer to chapter 2.6 where the meaning of the configuration adresses are explained.
2.3.1. Step 1
1. To enter into the configuration mode, switch the pump on
while keeping the keys "MODE" and START/STOP" pressed.
2.3.2. Step 2
If the PIN code was accepted the pump will show the display of chapter 2.3.1 again, but
you have now access to all writeable addresses in the list of chapter 2.6.
Select therefore any address in the left display (see next side):
14.164.A_A707en.SM.V1.01.doc
707 prog
5. Enter the PIN Code. If no PIN code was configured before,
default PIN 0 will be active. For PIN 0 press the key "10"
while keeping the key "1" pressed.
Cod
----
ARGUS Medical AG 10.02.04 AS
PUMP CONFIGURATIONS
2. Press key "START".
3. Press key "MODE".
4. Press key "START" again. The
3.
display will show "Cod ----".
Pre
6. Press key "START" to
acknowledge the en-
tered PIN code.
If the PIN was not
accepted, the display
will change back to
step 2.