5.5.3. Volume calibration with the pump integrated calibration program
Needed equipment:
1. Insert a new IV-set (only recommended IV-sets may be used, see appendix in
the user manual) in the pump (filled with water) and connect the infusion line
to a measuring cup on a balance.
2. Enter the configuration mode as described in chapter in chapter 2.3.
Enter the value "123" at address 399. This will force the pump to enter the
calibration mode which will be displayed as "bAL tArA".
3. Press the key "START/STOP". The pump delivers a volume of 5 ml at an infu-
sion rate of 250 ml/h ("tArA" is flashing).
4. When "tArA" stops flashing, reset the balance to 0.
5. Press the key "START/STOP". The pump will display "tM 215" (infusion de-
livery time is 215 seconds). It should now deliver a volume of 15 ml at an infu-
sion rate of 250 ml/h.
6. After the delivery time has elapsed, the pump stops and shows "baL." "12.75"
in the displays.
Enter now the value of the balance, e.g. 14.90 g. This value must be within the
range of 12.75 – 17.25. Otherwise switch the pump off and restart the volume
calibration.
7. Press the key "START/STOP" to acknowledge the entered value.
8. The pump displays the new correction factor, e.g. 1013.
9. Press the key "START/STOP" to store the new correction factor acknowledged
by a buzzer sound. The pump display changes back to "bAL tArA" again.
10. Switch the pump off. Do not remove the infusion set!
11. Perform a control measurement with an infusion rate of 250 ml/h and an infu-
sion total of 15 ml. Repeat the calibration procedure if necessary.

5.6. Pump specifications

Please refer to the user manual for the specifications (chapter 8).
14.164.A_A707en.SM.V1.01.doc
- balance with a resolution of 0,1g at least
ARGUS Medical AG 10.02.04 AS