5.7. Fault codes

A technical failure will be signalled by the pump with a continuous alarm display and a
continuous sound. During this state, the fault code which causes the pump to fail can be
displayed by pressing the key "MODE".
IF the pump was switched OFF after a detected failure, the fault code will be stored in the
configuration of the pump, please refer to chapter 2.6 (Addresses 380-389).
The possible fault codes registered in the configuration are listed in the table below:
Fault
Failure
Code
F_21
ROM test
F_22
ROM check (Runtime)
F_23
RAM test/check
F_24
XRAM test/check
F_25
CPU test
F_26
Invalid function menu
F_27
EEPROM data invalid
F_28
RTC data invalid, no RTC etc
F_29
Stepper motor power test (delayed 5s)
F_32
5Volt supply out of range
F_33
24Volt supply out of range (delayed 5s)
F_37
Downstream pressure sensor test failed (always > 4.7V, delayed 5s)
F_38
Upstream pressure sensor test failed (always > 4.7V, delayed 5s)
F_39
Downstream pressure sensor test failed (dynamical test failed)
F_40
Upstream pressure sensor test failed (dynamical test failed)
F_44
Address invalid for config-eeprom
F_45
Address invalid for history-eeprom
F_46
Frequency from uC or RTC out of range
F_47
Display-print not present
F_48
Key(s) too long active
F_49
Sensor-print not present
F_50
AIL (Air in line) detector test failed
F_51
Movement test failed (Home-Pulse < (Hall / Home))
F_52
Movement test failed (Home-Pulse > (Hall / Home))
F_53
More than one rotation at 'STOP' without 'KVO'
F_54
Infused sum <> Calculated sum (Rotations)
F_55
Frequency calculation
F_57
Rotation (SW overflow)
F_58
Volsum control
We recommend replacing the main board in case a fault code is not included in this list
above.