5. MAINTENANCE

5.1. General

CAUTION!
Only authorized persons who have been trained by ARGUS Medical AG or by the local
distributor are allowed to service the ARGUS 707 infusion pump. In case of repair request,
send the unit to the local distributor with a report outlining the exact nature of the failure.
More information is available from:
CAUTION!
The safety standard check has to be performed at least every 24 month or after
10'000 hours of operation. The check has to be done in accordance to the chapter 7.
No special maintenance of the ARGUS 707 infusion pump is necessary. There are no
wear and tear parts.

5.2. Recalibration

5.2.1. General
Unless otherwise specified by the customer, the ARGUS 707 has been calibrated by the
manufacturer with the CODAN L86 infusion set. If a different infusion set is used (see
recommended list in the appendix of the user manual), a recalibration is required.
CAUTION!
A recalibration with a different IV-set always requires a recalibration of both pres-
sure sensors and a volumetric calibration!
CAUTION!
It is mandatory to execute first the calibration procedure of the pressure sensors and
afterwards the volume calibration.

5.3. Pressure calibration

5.3.1. General
The volumetric infusion pump ARGUS 707 contains two pressure sensors:
- One upstream sensor, bottle side (left input)
- One downstream sensor, patient side (right output).
CAUTION!
A pressure calibration becomes necessary if the pressure control measurement
was not accurate enough, a new IV-set configured or a pressure sensor re-
placed or the main board is replaced.
Needed equipment: - a manometer with a resolution of 0,05 bar.
14.164.A_A707en.SM.V1.01.doc
ARGUS Medical AG
CH-3627 Heimberg / Switzerland
E-mail: info@argusmedical.com
- a 3-way stop cock
- the chosen IV-set (refer to the list
including the recommended IV-set in the user manual)
ARGUS Medical AG 10.02.04 AS
MAINTENANCE