10. APPENDIX

Safety standard check (SSC)
Serial-no.: ...................................................
Hospital:
Department:
Customer:
The SSC has to be performed at least every 24 months or after 10'000 hours of operation. The check has to be
done in accordance to the user- and service manuals.
1
Visual check for damage, cleanness and completeness
Test the function of the stop flow clamp
2
3
Keep "MODE" pressed while switching on the pump
Connect/Disconnect the pump to the mains
4
Test the drop detector by simulating drops
5
6
Test the door switch, open and close the door
7
Open the door and remove any IV-set
8
Install a water filled IV-set, close the door
Set rate to 333.3 m/h, press "START", disconnect the
9
drop detector
Press "MODE"
10
11
Check the external connector "nurse call"
12
Calibration of the pressure sensors
IV-set type used:
Codan......... Other.......................................
13
Volume calibration
14
Charge the battery while the pump is running 16
hours, at a rate of 30.0 ml/h
15
Battery check at a rate of 30 ml/h.
Run the battery test until the pump switches
off automatically
16
Printout the pump history (refer to chapter 3.4)
17
Electrical test according to EN60601-1
(all measurements made with a power cable 2.5 m)
18
Charge the battery after this test!
The pump has passed the SSC and is safe for use
Date:
Name:
Signature:
14.164.A_A707en.SM.V1.01.doc
ARGUS 707
....................................................................................................
....................................................................................................
....................................................................................................
....................................................................................................
....................................................................................................
....................................................................................................
ARGUS Medical AG 10.02.04 AS
ARGUS Medical AG
Inventory-no:..............................................
- Housing, labels, accessories, connectors, power
cable, etc.
- Proper movement of the clamp
Display pump type and software release
-
Display of 2, 4, 7, F., in numeric display
-
Display of all operation- and alarm indicators
-
The indicator "external supply" lights up
-
Check the green "drop" indicator
-
Door open the "door" indicator lights up
-
Door close the "door" indicator turns off
-
Indicator "air bubble" lights up
-
Indicator "air bubble" turns off
-
The red indicator "drop" lights up (delayed)
-
The acoustical alarm turns on
-
The acoustical alarm mutes
-
Relay contact switches (see chapter 9)
-
See chapter 5.3 and 5.4 of the service manual
-
See chapter 5.5 of the service manual
-
The indicator "external supply" must light
-
The green indicator "battery" lights up during this
-
test
Check the battery run time by checking the latest
-
history entries, > 4h 30min
Check the pump internal time and date
-
Measurements attached
-
APPENDIX
38 / 38