Specifications Of Accuracy - Iwatsu SS-530 Instruction Manual

Hide thumbs Also See for SS-530:
Table of Contents

Advertisement

Specifications of accuracy

Accuracy warranty conditions
Amplitude accuracy*
*3. Unless there is any change in the state of the facing core
surfaces, which includes scratches, adhesion of foreign objects,
or any change in the operating environment.
Specifications of Input, Output, and Measurement
• Accuracy warranty period
1 year (until the upper jaw has been retracted
and locked up to 10,000 cycles)
• Guaranteed accuracy period after adjustment
made by IWATSU
1 year
• Temperature and humidity for guaranteed
accuracy
23°C±5°C (73°F±9°F), 80% RH or less
• Warm-up time
At least 30 minutes
• Power voltage range
±12 V DC ±0.5 V
3
30 A range: ±3.0% rdg. ±1 mV
(Typical: ±1.0% rdg. ±1 mV
[for 10 A rms or less])
5 A range: ±3.0% rdg. ±1 mV
(Typical: ±1.0% rdg. ±1 mV)
0.5 A range: ±3.0% rdg. ±10 mV
(Typical: ±1.0% rdg. ±10 mV)
For a direct current and a sine-wave current
with frequencies of 45 Hz to 66 Hz within the
maximum peak current value of each current
range
3
73

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ss-531