Measuring Currents

When the device has entered protection mode

To protect the device against self-generated heat, it enters
protection mode when the temperature of the junction box
exceeds a specified level.
Green lights up
Reds blink rapidly
Orange blinks rapidly
Greens blink rapidly
In protection mode, the device cannot correctly measure any
current. Moreover, you cannot switch the current ranges.
When the device has entered protection mode, follow the
procedure presented on the following pages to restore it to
normal operation.
If the device has entered protection mode, it is
recommended to re-calibrate it because internal components
may have been subjected to thermal stress.