Product specifications

Type:
Measuring range:
Resolution:
Laboratory Accuracy:
Display:
Acoustic:
Operating temperature:
Automatic Switch-off:
Weight:
Long term storage ranges
Storage/transport
temperature:
Humidity:
Battery:
Pressure:
This infrared thermometer meets requirements established in ASTM Standard
E 1965-98 for the thermometer system. Full responsibility for the conformance
of the product to the standard is assumed by Kaz USA, Inc., a Helen of Troy
Company, 400 Donald Lynch Blvd., Marlborough, MA 01752, USA.
ASTM laboratory accuracy requirements for the thermometer only (not including
a lens filter) in the display range of 36 °C to 39 °C (96.8 °F to 102.2 °F) for infrared
thermometers is ± 0.2 °C (± 0.4 °F), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is
±0.1 °C (±0.2 °F).
This device conforms to the following standards:
IEC 60601-1:2014 - Medical electrical equipment. General requirements for basic
safety and essential performance
ISO 80601-2-56:2009 - Particular requirements for basic safety and essential
performance of clinical thermometers for body temperature measurement
ASTM E1965-98:2009 - Standard Specification for Infrared Thermometers for
Intermittent Determination of Patient Temperature
12
No touch + forehead thermometer (NTF3000CAV1)
34.4 °C – 42.2 °C (93.9 °F – 108 °F)
0.1 °C (0.1 °F)
±0.2 °C for the range 35 °C – 42 °C
(±0.4 °F for the range 95 °F – 107.6 °F)
Ambient Temperature: 15 °C to 40 °C (59 °F to 104 °F)
Liquid Crystal Display, 4 digits plus special icons
Audio:
Normal temp range = single long beep for 2 sec duration
Fever = Red or Yellow temp range: 10 short beeps
15 °C – 40 °C (59 °F – 104 °F)
Approximately 60 seconds after last measurement
has been taken
99.5 g (with batteries), 77.1 g (w/o batteries)
-25 °C to 60 °C (-13 °F – 140 °F)
15–95% non condensing
(2) AA Batteries - at least 1000 measurements
700-1060hPA (0.7-1.06 atm)