Fig; Legend Of Fig; Item Designation; Lock Bolt - Braun Infusomat Space Service Manual

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

2

Fig.: 3 - 25

Legend of fig. 3 - 25:

Item Designation

1
Screw DELTA PT 30x8 WN 5452 TORX plus 10IP
2
Pump drive motor
EN
Infusomat® Space 6.0
4. Unscrew one screw
5. Move the lock bolt drive by hand in the middle position.
6. Unscrew two screws
Note
The lock bolt drive is slightly engaged when installed. When dis-
mantling the drive make sure that the toothed wheels do not rub
along the metal edges of the inner frame as the toothed wheels
can easily be damaged. The toothed wheels are very easily dam-
aged.
1
7. Lift the claw on the right side and push it to the left until the
8. Lift the lock bolt over the guide rivet of the inner frame.
Disassembly / Assembly
(Fig.: 3 - 25 / Item
motor
(Fig.: 3 - 25 / Item
2) clockwise out of the holder and
remove it from the pump frame.
(Fig.: 3 - 24 / Item
bolt drive
(Fig.: 3 - 24 / Item
(Fig.: 3 - 24 / Item
1).
guide rivet is exactly positioned in the mounting hole of the

lock bolt.

3
1), turn the pump drive
2) and pull the lock
4) off the lock bolt tappet