3
Disassembly / Assembly

3.3 Cover for Drop Sensor Connector

Fig.: 3 - 9
Legend of fig. 3 - 9:
Item Designation
1
Cover for drop sensor connector

3.4 Emergency Release Plug

Fig.: 3 - 10
Legend of fig. 3 - 10:
Item Designation
1
Emergency Release Plug
3 - 10
Designation
Cover, drop sensor connector . . . . . . . . . . . . . . . . . . . . 3452 1577
Disassembly
1
1. Carefully remove the cover
sensor connector from the rear to the front and out of the
housing.
Note
Be careful not to damage the lugs in the housing bottom and up-
per part.
Designation
Emergency release plug ISP incl. O-ring
(see „Service Parts and Screw Kit" ➨ pg. 3 - 7)
Disassembly
1. Turn the emergency release plug
a spanner, e.g. the emergency release crank in the battery
compartment cover, through 90° to the right and pull the
plug out of the housing.
Note
The spanner must be pressed carefully until stop in the opening of
the emergency release plug. The hexagon socket of the emergency
release plug tapers to the inside to guarantee a certain bond when
1
the plug is removed.
Order No.