Inner Frame And Housing Front Panel - Braun Infusomat Space Service Manual

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

3.9 Inner Frame and Housing Front Panel

2
1
7
Fig.: 3 - 22
Legend of fig. 3 - 22:
Item Designation
1
Connector for ribbon cable, operating unit
2
Lock bolt drive connector
3
Pressure sensor connector, downstream
4
Air inline sensor connector
5
Safety clamp connector
6
Pump drive connector
7
Pressure sensor connector, upstream
EN
Infusomat® Space 6.0
3
4
5
6
Disassembly / Assembly
Designation
Screws
(see „Service Parts and Screw Kit" ➨ pg. 3 - 7)
Note
Please pay attention to the corresponding notes during assembly
and installation
(see „Inner Frame and Housing Front Panel" ➨
pg. 3 -
44).
Disassembly
Note
When dismounting the inner frame and the housing front panel
please pay attention to the length of the connection cables to the
processor PCB.
1. Fold the locking of the ribbon cable connector
Item
4) to the air inline sensor up and pull the ribbon cable
off the connector.
2. Pull the safety clamp connector
processor PCB.
3. Pull the pump drive connector
processor PCB.
3
Order No.
(Fig.: 3 - 22 /
(Fig.: 3 - 22 / Item
5) off the
(Fig.: 3 - 22 / Item
6) off the
3 - 23

Hide quick links:

Advertisement

Table of Contents
loading