Infusomat® Space 6.0

Unit Diagnosis / Calibration

2 -
The alarms of the Infusomat® Space are classified in 5 categories.
These categories are listed hereafter according to their impor-
tance.
-
Alarm advice
In case of unacceptable inputs corresponding messages are
displayed (e.g. "Caution! Rate out of range", "The parameter
cannot be changed") and a beep sounds.
-
Pre-alarm
Pre-alarms are triggered several minutes (depending on the
service settings) before the operating alarms.
-
Reminder alarm
A reminder alarm is triggered if the device is not operated for
two minutes when input or operation was not finished.
-
Operating alarm
In case of an operating alarm the infusion is stopped. An au-
dible signal is released, the red LED flashes and a staff call is
triggered. The message "Alarm" and the cause of the alarm
appear on the display.
-
Device alarm
The most important alarms and error codes as well as their mean-
ing and possible fault clearance are specified in the following
lists.
2
2 - 1