Housing Bottom Part - Braun Infusomat Space Service Manual

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

3
Disassembly / Assembly

3.8 Housing Bottom Part

6
Fig.: 3 - 21
Legend of fig. 3 - 21:
Item Designation
1
Snap-in hook
2
Loudspeaker
3
Housing bottom part
4
Screw DELTA PT 30x12 WN 5452 A2 TORX plus 10IP
5
Cover cap
6
Screw DELTA PT 30x8 WN 5452 A2 TORX plus 10IP
3 - 22
1
2
3
4
5
Designation
Loudspeaker SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3452 0937
Housing bottom part ISP . . . . . . . . . . . . . . . . . . . . . . . 3452 1305
Rating plate ISP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . On request
Snap-in hook SP with leaf spring
(see „Service Parts and Screw Kit" ➨ pg. 3 - 7)
Screws and cover caps
(see „Service Parts and Screw Kit" ➨ pg. 3 - 7)
Note
Please pay attention to the corresponding notes during assembly
and installation
(see „Housing Bottom Part" ➨ pg. 3 -
Disassembly
1. Remove the drop sensor connector (not depicted) from the
housing bottom part
(Fig.: 3 - 21 / Item
Note
The drop sensor connector is integrated in the processor PCB.
2. Pierce four cover caps
(Fig.: 3 - 21 / Item
screwdriver and pry the cover caps out.
3. Unscrew three screws
(Fig.: 3 - 21 / Item
(Fig.: 3 - 21 / Item
4).
4. Take off the housing bottom part.
5. Pull the loudspeaker connector off
processor PCB and remove the loudspeaker.
6. Pull the snap-in hook
(Fig.: 3 - 21 / Item
bottom part.
Order No.
45).
3).
5) with a small
6) and one screw
(Fig.: 3 - 21 / Item
2) the
1) out of the housing
Infusomat® Space 6.0