Disassembly / Assembly
1. Make sure that the following components are sufficiently lu-
bricated when installing the inner frame. If the grease film is
not sufficient (stiff or jerky movements) grease these areas
slightly with Polylub GLY 501.
Note
If chemical substances such as lubricating grease are used the
safety data sheets are to be observed
pg. 0 -
7).
-
Plastic guides
(Fig.: 3 - 44 / Item
-
Spring bearings and dovetail of the spring holder
(Fig.: 3 - 44 / Item
-
Guide (6 mm straight pin) of the pump frame at the bear-
ings
(Fig.: 3 - 44 / Item
-
Toothed wheels
(Fig.: 3 - 44 / Item
-
Movement areas
(Fig.: 3 - 44 / Item
frame
-
Toothed wheels
(Fig.: 3 - 44 / Item
drive
-
Guides
(Fig.: 3 - 44 / Item
lock bolt
2. Pay attention to the operating pin of the linear potentiometer
when installing the lock bolt drive. The pin must engage in
the fork for the potentiometer when the spindle tappet is
clipped in.
3. Check all toothed wheels for damage after installation of the

lock bolt drive.

4. The spring holder of the pressure adjustment unit must run
smoothly with its dovetail guide in the guide of the pump
frame. Check smooth running without springs inserted before
final assembly.
(see „Safety data sheets" ➨
1) of the lock bolt
5)
6)
7) of the pump drive
3) of the pump
4) of the lock bolt
8) and supporting area of the