1. On the Counter Navigation screen, push SYSTEM.
2. Select the Units and Alarms tab.
3. Select the HMP RH/T check box to activate or deactivate the HMP probe.
Manage backup and restore settings
Users with administrator or System Settings permissions can make a backup of configurable
settings, and restore the backup settings as needed.
Make a backup of configurable settings
1. On the Counter Navigation screen, push SYSTEM.
2. Select the Options and Accessories tab.
3. Push BACK-UP.
An electronic copy of the current configuration is stored in instrument memory. This version can
be restored with the Restore function.
Restore settings from backup
Use the Restore function when settings become corrupt, incorrectly modified or when the software
performs abnormally.
1. On the Counter Navigation screen, push SYSTEM.
2. Select the Options and Accessories tab.
3. Push RESTORE.
4. Push RESTART.
Configurations
Specific configurations can be saved as a backup and/or copied as necessary between particle
counters.
Copy a configuration
To copy a configuration:
1. On the Counter Navigation screen, push SYSTEM.
2. Select the Configuration tab.
3. Insert a USB drive into the USB host connector. Refer to
4. Push COPY CONFIGURATION TO USB.
5. A confirmation message will show. Push OK.
6. Remove the USB drive.
Install a configuration
To install a configuration from a USB drive:
1. On the Counter Navigation screen, push SYSTEM.
2. Select the Configuration tab.
3. Insert the USB drive with the configuration data into the USB host connector. Refer to
on page 12.
4. Push READ CONFIGURATION FROM USB.
5. A confirmation message will show.
6. Remove the USB drive.
7. Push OK to restart the instrument and load the new configuration.
24 English
Figure 3
on page 12.
Figure 3