Installing the ON/OFF assembly
Fig. 1

Preparations

NOTE
Before installing the new ON/OFF assembly, it must be set to the local mains voltage
according to the circuit diagram (see Fig. 4).
Check the fuse values of fuses F1 and F2.
For mains voltages of 200 V AC to 240 V AC, 15 A slow-blow fuses must be used.
For mains voltages of 100 V AC to 127 V AC, 20 A slow-blow fuses must be used.
Siemens AG
Medical Solutions
The ON/OFF assembly is set to 230 V AC by default. If the local
mains voltage differs from that, the ON/OFF assembly must be set
to the local mains voltage prior to startup. Please observe the
fuse values for fuses F1 and F2 for the relevant mains voltage.
n.a.
SP00-000.841.01
Rev. 01
Fig. 2
Page 1 of 10
08.05
CS PS 24
3 - 1
3
3
ARCADIS
System Manual