Sony LMD-X3200MD Manual page 21

Hide thumbs Also See for LMD-X3200MD:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 18
Important EMC notices for use in medical environments
• The LMD-X3200MD needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in the instructions for use.
• The LMD-X3200MD is intended for use in a professional healthcare facility environment.
• Portable and mobile RF communications equipment, such as cellular phones, can affect the LMD-
X3200MD.
Warning
• Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of
the LMD-X3200MD. Otherwise, degradation of the performance of this equipment could result.
• If the LMD-X3200MD will be used adjacent to or stacked with other equipment, normal operation of the
LMD-X3200MD under such configurations should be verified via observation.
• The use of accessories and cables other than those specified, with the exception of replacement parts
sold by Sony Corporation, may result in increased emissions or decreased immunity of the LMD-
X3200MD.
Guidance and manufacturer’s declaration – electromagnetic emissions
The LMD-X3200MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the LMD-X3200MD should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
CISPR 32
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
The LMD-X3200MD uses RF energy only for its
internal function. Therefore, its RF emissions are
Group 1
very low and are not likely to cause any
interference in nearby electronic equipment.
The LMD-X3200MD is suitable for use in all
establishments, including domestic
Class B
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Class D
(AC input)
Not applicable
(DC input)
Complies
(AC input)
Not applicable
(DC input)
Electromagnetic environment – guidance
5