For customers in Canada

This unit has been certified according to Standard
CAN/CSA-C22.2 No. 60601-1.
Important safeguards and notices for use
in the medical environments
1. All devices connected to the unit must be
certified or compliant according to IEC 60601-1,
IEC 60950-1, and IEC 60065 standards and
other IEC/ISO standards applicable to the
devices.
2. Furthermore, the system as a whole must
comply with IEC 60601-1 standards. All
peripheral devices connected to the signal
input/output sections of the unit constitute the
medical-use system, and therefore, the user is
responsible for ensuring that the system as a
whole complies with IEC 60601-1 standards. If
in doubt, consult qualified Sony service
personnel.
3. Connecting the unit to other devices may
increase the leakage current.
4. For all peripheral devices connected to the unit
that operate on commercial power supplies
and do not comply with IEC 60601-1 standards,
incorporate an isolation transformer that
complies with IEC 60601-1 standards and
connect to the commercial power supply via
the transformer.
5. The unit generates, uses, and may radiate radio
frequency energy. If it is not installed and used
in accordance with the instruction manual, it
may cause interference on other devices. If the
unit causes interference (which can be
determined by disconnecting the power cord
from the unit), try the following.
– Relocate the unit with respect to the affected
devices.
– Connect the unit and the affected devices to
different branch circuits.
For more information, consult qualified Sony
service personnel.
(Applicable standard: IEC 60601-1-2)