Disable the PC network adapter temporarily:
<Start> <Settings> <Control panel> <System> <Hardware> <Device Manager> <Network adapters>
Switch on/off the scanner's power.
Software is not to be launched during the initialization time of the scanner (about 120 s).
Use the network administration utility Ping to test the reachability of the scanner. Call the Windows <Command
Prompt> and execute ping 192.168.42.30 (exemplary IP address)
Use the network protocol Telnet to provide a bidirectional text-oriented communication to the scanner by using
the scanner specific commands. Call the Windows <Command Prompt> and execute telnet 192.168.42.30 2727
(exemplary IP address and port).
Note: Telnet is not available on all Windows computers by default!
Verify correct C
port. If existing use another C
OM
To avoid communication problems the properties of all used C
<Start> <Settings> <Control Panel> <System> <Hardware> <Device Manager> <Ports (COM & LPT)>
<Communications Port COMx> <Ports Settings> <Advanced> <Receive Buffer>: Low
In case of using plug-in serial cards, disable <CTS/RTS> for the Auto Flow Control.
<Start> <Settings> <Control Panel> <System> <Hardware> <Device Manager> <Ports (COM & LPT)>
<Communications Port COMx> <Ports Settings> <Advanced>
Avoid the use of third party USB/RS232 converters!
Ensuring the RS232/485 converter is close to the PC (and not to the scanner)
In case of communication errors, the baud rate should be reduced step by step.
Software is not to be launched during the initialization time of the scanner (about 120 s).
On some computers, the performance of the software can be increased dramatically by switching off the
hardware graphic acceleration:
<Start> <Settings> <Control Panel> <Display> <Settings> <Advanced> <Troubleshooting> <Hardware
acceleration>: None
A too high display resolution can cause problems. Test temporarily a setting of 800x600 pixel at high color (16
bit).
For several reasons the PC can be overloaded. This can be checked by running the scanner software and
pushing CTRL+ALT+DEL at the same time. In the task manager a window for the system performance can be
selected. The value must be every time much below 100%.
In some cases, the graphic card of the computer can interrupt serial communications for too long a time. S3-
graphic cards are known to exhibit this problem. The only solution is to exchange the S3-graphic card.
Possible Cause / Solution
port on the computer.