Pace PRC 2000 Service Manual page 7

Hide thumbs Also See for PRC 2000:
Table of Contents

Advertisement

GENERAL INFORMATION
THERMAL MANAGEMENT CENTER
VACUUM AND AIR
Measurements at Front Panel SNAP-VAC and PRESSURE Ports of power source.
Vacuum Rise Time :
Vacuum:
Pressure:
Air Flow:
HANDPIECES
Set Temperature Range
of SensaTemp Handpieces:
Digital Readout Resolution: ± 1° (°C or °F)
Tip Temperature Stability:
True minimum and maximum Operating Tip Temperatures may vary
depending on handpiece & tip selection.
EOS/ESD
Tip-To-Ground
Resistance:
AC Leakage :
PIK AND PASTE
Vacuum
(at PIK-VAC Port):
Pressure
(at PASTE DISP Port):
2
Evacuates 33 cc (2 cubic inch) volume
to 25 cm Hg. (10 in. Hg.) in 150 ms.
51 cm Hg. (20 in. Hg.) (nominal)
.48 Bar (7 P.S.I.) (nominal, "MAX" setting)
13 SLPM (0.46 SCFM) maximum
38°C to 482°C (100°F to 900°F) (see note).
± 1.1°C (± 2°F) at idle from Set Tip Temperature.
NOTE
Less than 5 ohms.
Less than 2 millivolts RMS from 50Hz to 500Hz.
7.6 cm Hg. (3 in. HG.) min.
2.41 Bar (35 P.S.I.) min.

Advertisement

Table of Contents