Dflt Factory Defaults Command; Dispfld; Displayed Field Command; Displayed Field Query - Omega CYD211 User Manual

Single input temperature monitor
Table of Contents

Advertisement

DFLT
Factory Defaults Command
Input:
DFLT 99[term]
Remarks:
Sets all configuration values to factory defaults and resets
the instrument. The "99" is included to prevent accidentally
setting the unit to defaults.

DISPFLD

Displayed Field Command

Input:
DISPFLD <source>[term]
Format:
n
<source> Specifies input data to display. Valid entries:
Example:
DISPFLD 1[term] – Displays Kelvin reading for the input.
DISPFLD?

Displayed Field Query

Input:
DISPFLD?[term]
Returned: <source>[term]
Format:
n
FRDG?
Fahrenheit Reading Query
Input:
FRDG?[term]
Returned: <temp value>[term]
Format:
±nnnnnn
Remarks:
Also see the RDGST? command.
INCRV

Input Curve Number Command

Input:
INCRV <curve number>[term]
Format:
nn
<curve number> Specifies which curve the input uses. If
Remarks:
Specifies curve the input uses for temperature conversion.
Example:
INCRV 21[term] – The input User Curve 21 for
temperature conversion.
Remote Operation
Omega Model CYD211 User's Manual
0 = Kelvin, 1 = Celsius, 2 = sensor units, 3 =
Fahrenheit.
(Refer to command for description)
specified curve parameters do not match
the input, the curve number defaults to 0.
Valid entries: 0 = none, 1 – 20 = standard
curves, 21 = user curve.
4-19