Fig. 2:
V
With the housing covers open and safety covers removed:
Back:
Left:
Right:
MOBILETT XP Digital
Locations with dangerous voltage
CAUTION: DC voltage (440V) at capacitor bank (D927)
CAUTION: DC voltage (300 V) from the battery block to PCB D982!
Always disconnect the battery plug from BK1-BK4.
CAUTION: DC voltage (440V) directly at the capacitor bank!
CAUTION: DC voltage (440V) directly at the capacitor bank!
CAUTION: AC voltage (> 100V or > 60V half line voltage) at power supplies
U1 and U2 as long as the power cord is connected.
SPR8-230.831.30.03.02
11.05
Page 10 of 74
CS SD 24