Inspection and maintenance
3.12
Patient leakage current measurement
SIE
Patient leakage current measurement
Observe the patient leakage current measurement information in these instructions.
•
Perform the patient leakage current measurement with all covers closed, according to
ARTD-002.731.17....
•
Measure the patient leakage current at the flat detector.
•
The limit values of 0.01 mA for direct current and 0.1 mA for alternating current may not
be exceeded.
•
The system must be switched on during measurement. Measuring devices with auto-
mated measuring sequences must therefore be set to manual measurement.
•
Document the measuring procedure and the measuring device used (designation and
serial number).
•
The value determined must be entered in the patient leakage current report and evalu-
ated.
NOTE
NOTE
Siemens AG
Medical Solutions
File the patient leakage current report in the "Certificates" register
no. 9 in the system binder.
A new report must be created if the patient leakage current mea-
surements are not documented.
A new patient leakage current report can be found in the startup in-
structions SPR8-230.814.30.03. Separate this report from these in-
structions, fill it out, and file it in the "Certificates" register no. 9 in
the system binder. If values are newly determined, they must be re-
corded as first measured values. The evaluation is omitted in this
case.