Replacing The Bottom Housing; Replacing The Fuses - Invacare Softform Active 2 Rx Service Manual

Powered hybrid support surface
Hide thumbs Also See for Softform Active 2 Rx:
Table of Contents

Advertisement

4.10

Replacing the Bottom Housing

R
efers to spare parts kit SP1633246
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
PH1 Phillips key
1.
Switch off the control unit and disconnect from
mains supply.
2. Remove the bottom housing (see 4.1).
3. Install the new bottom housing.
4. Position and tighten the provided screws.

4.11 Replacing the Fuses

R
efers to spare parts kit SP1633143
PH1 Phillips key
1.
Switch off the control unit and disconnect from
mains supply.
2. Turn the unit upside down.
3. Unscrew the fuse holders A. Remove them from
the unit.
4. Remove the fuses from their holders. Discard
the defective fuses.
5. Install the new fuses on their holders.
6. Position the fuse holders back to the unit and
tighten them.
60129676-A
A
A
Fig. 4-30
Fig. 4-31
Invacare
®
Softform Active
®
2 Rx
17

Advertisement

Table of Contents
loading

Table of Contents