4.8 Replacing the Motor

R
efers to spare parts kit SP1633126
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
PH1 Phillips key
1.
Switch off the control unit and disconnect from
mains supply.
2. Remove the bottom housing (see 4.1).
3. Unplug the three silicon tubes A connected to
the motor.
4. Unplug the black B and green C wires from the
microswitch.
Risk of malfunction or damage to the
Control Unit
Ensure the black B and green C wires are
plugged back into their original positions.
5. Loosen and remove the screws D fixing the motor
to the front housing.
6. Discard the non-functioning motor.
7. Install the new motor on its position.
8. Position and tighten the provided screws.
9. Plug tubes A and the wires to the new motor.
10. Reinstall the bottom housing.