15. Pick the longer tube connected to the CPR
connector B, put it around the motor, clockwise,
and plug it back to the connector, threading it
under the microswitch wires D.
16. Install the CPR connector in its place, making
sure the longer silicon tube is underneath.
17. Slide the T-valve underneath the microswitch,
connect the wider tube to the compressor, and
the thinner one to the user interface, making sure
it goes under all the other tubes.
18. Reinstall the bottom housing.
60129676-A
D
D
Fig. 4-21
Fig. 4-22
Fig. 4-23
Invacare
®
Softform Active
®
2 Rx
B
13