Guidance and manufacturer's declaration - electromagnetic immunity
OMRON BPM including AC-adapter is intended for use in the electromagnetic environment specified
below. The customer or the user of this OMRON BPM including AC-adapter should assure that it is
used in such environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply
inputlines
IEC 61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
Note: U
is the A.C. mains voltage prior to application of the test level.
T
IEC 60601 test
Compliance level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV line(s) to
±1 kV line(s) to
line(s)
line(s)
±2 kV line(s) to
±2 kV line(s) to
earth
earth
<5 % U
(>95 %
<5 % U
T
dip in U
)
dip in U
T
for 0.5 cycle
for 0.5 cycle
40 % U
(60 % dip
40 % U
T
in U
)
in U
T
for 5 cycles
for 5 cycles
70 % U
(30 % dip
70 % U
T
in U
)
in U
T
for 25 cycles
for 25 cycles
<5 % U
(>95 %
<5 % U
T
dip in U
)
dip in U
T
for 5 sec.
for 5 sec.
3 A/m
3 A/m
Electromagnetic environment -
Floor should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of
a typical commercial and/or hospital
environment.
Mains power quality should be that of
a typical commercial and/or hospital
environment.
(>95 %
T
)
T
Mains power quality should be that of
a typical commercial and/or hospital
(60 % dip
T
environment. If the user of the
)
T
OMRON BPM including AC-adapter
requires continued operation during
(30 % dip
T
power mains interruption, it is
)
T
recommended that the OMRON BPM
including AC-adapter be powered
from an uninterruptible power supply.
(>95 %
T
)
T
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
37
guidance