Omron IntelliSense HEM-907XL Instruction Manual

Omron IntelliSense HEM-907XL Instruction Manual

Omron healthcare instruction manual blood pressure monitor hem-907xl
Hide thumbs Also See for IntelliSense HEM-907XL:

Advertisement

INSTRUCTION MANUAL
IntelliSense™ Blood Pressure Monitor
HEM-907XL
Model
Thank you for purchasing the OMRON
IntelliSense™ Blood Pressure Monitor.
Please thoroughly read this Instruction Manual
before using this monitor to ensure safe and
accurate use.
Please keep this manual near the monitor all
the time for future reference.

Table of Contents

Be Sure to Read This Section
Exemptions ........................................2
Notes on Safety .................................3
Know Your Unit
Features of the Product .....................7
Components of the Product ...............8
Options ...........................................9
Features and Functions ...................10
Names of the Parts ......................10
while the unit is in use ..................12
Preparations before Measurement
How to Apply the Cuff ......................14
of Battery Pack .............................17
SINGLE Mode ..............................20
AVG Mode ....................................22
MANU Mode .................................24
List of Error Codes ...........................28
Troubleshooting................................29
Specifications
Caution.............................................30
Specifications ...................................31
16
............

Advertisement

Table of Contents
loading

Summary of Contents for Omron IntelliSense HEM-907XL

  • Page 1: Table Of Contents

    INSTRUCTION MANUAL IntelliSense™ Blood Pressure Monitor HEM-907XL Model Thank you for purchasing the OMRON IntelliSense™ Blood Pressure Monitor. Please thoroughly read this Instruction Manual before using this monitor to ensure safe and accurate use. Please keep this manual near the monitor all the time for future reference.
  • Page 2: Exemptions

    EXEMPTIONS OMRON does not accept liability and warranty becomes void under the following circumstances: 1. When persons, not authorized by OMRON, perform repairs or modifications of this product. 2. When use and/or operation of this device is adversely effected by a device not manufactured by OMRON.
  • Page 3 NOTES ON SAFETY NOTES ON SAFETY Unplug the AC adapter from the electric outlet if this unit is unused for an extended period of time. Unplug the AC adapter from the electric outlet when installing, removing, or cleaning the unit. Confirm readings with a stethoscope when an irregular pulse wave is displayed or when the measured value is questionable or erratic.
  • Page 4: Features Of The Product

    (3) Display "Out of use" on this unit so that it cannot be used. (4) Contact Omron for repair service (1-877-216-1336). FEATURES OF THE PRODUCT OMRON IntelliSense™ Blood Pressure unit, Model HEM-907XL is developed to measure blood pressure and pulse rate accurately and simply in a doctor's office, examination room, or patient bedside.
  • Page 5: Components Of The Product

    Accessories (Included and also available separately) Cuff / bladder set Extra Large (with built-in bladder) * Model: HEM-907-CX19 For use with the HEM-907XL unit only. Applicable arm circumference: 17" to 20" ( 42 to 50 cm) AC adapter Air tube Battery pack 1.0m (39 1/3")
  • Page 6: Features And Functions

    FEATURES AND FUNCTIONS Names of the Parts Main unit 1. Display (See Page 11 for detail) 2. HIDE (non-display) Button 3. DC jack 4. P-SET (pressure setting) Knob 5. MODE Selector 1. Display: Displays blood pressure and, pulse rate readings, and oscillation pulse level. 2.
  • Page 7: Functions And Setting

    FEATURES AND FUNCTIONS Functions setting (1) Inflation level setting P - SET AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected. The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure (approximately 30-40 mmHg above the patient’s systolic pressure).
  • Page 8: How To Apply The Cuff

    HOW TO APPLY THE CUFF The cuff of OMRON IntelliSense™ Blood Pressure Monitor HEM-907XL plays an important role of collecting the information on blood vessels. Please wrap the cuff according to the procedure, below. Do not wrap the cuff over an arm to which intravenous injection or transfusion is being conducted, or when otherwise contraindicated.
  • Page 9: How To Use The Power Source

    HOW TO USE THE POWER SOURCE How to use the AC adapter Do not use this unit in the presence of Be sure to use the AC adapter from the flammable gas, or anesthetics, or in a high power supply of 110 VAC. pressure oxygen room or oxygen tent.
  • Page 10: How To Measure Blood Pressure

    HOW TO USE THE POWER SOURCE Battery life • You can use the unit for approximately three hundred measurements with one charge. • Approximate life of battery is two years. However the battery life from each charging may be shortened depending on the state of using. If the interval between charging becomes short and icon appears frequently, replace it.
  • Page 11 HOW TO MEASURE BLOOD PRESSURE (IN SINGLE MODE) 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to "SINGLE". 3. Set the P-SET (inflation level) Knob to "AUTO" or to the target pressure value.
  • Page 12 HOW TO MEASURE BLOOD PRESSURE (IN AVERAGE MODE) 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to "AVG". The factory-set values are set as follows: • Number of measurements: 2 • Waiting time until the 1st measurement: 0 sec. •...
  • Page 13 HOW TO MEASURE BLOOD PRESSURE (IN MANUAL MODE) 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to "MANU". 3. Set the P-SET (inflation level setting) Knob to "AUTO" or the target pressure value. 4.
  • Page 14: How To Check Pressure Accuracy

    HOW TO CHECK PRESSURE ACCURACY Accuracy of pressure display can be checked in the CHECK Mode. What you need to prepare (1) Calibrated mercury manometer (including inflation bulb), (2) T-tube, (3) two air tubes, and (4) a sturdy cylindrical shaped object on which the cuff is wrapped. How to check 1.
  • Page 15: List Of Error Codes

    LIST OF ERROR CODES Error code Explanation Inflation error • Confirm that the air tube connecting the cuff and • When the pressure does not exceed 12 the main unit is connected securely. mmHg within the set time after the start of inflation •...
  • Page 16: Caution

    FIVE YEAR LIMITED WARRANTY Your HEM-907XL IntelliSense™ Automatic Blood Pressure Monitor is warranted to be free from manufacturing defects for a period of five years under normal use. The five year warranty excludes the monitor cuff. The cuff is warranted for a one year period. This warranty extends only to the original retail purchaser.
  • Page 17 For Customer Service Call Toll Free 1-877-216-1336 Manufactured by OMRON Distributed by: OMRON HEALTHCARE, INC. 300 Lakeview Parkway Vernon Hills, Illinois 60061 www.omronhealthcare.com Copyright © 2003 OMRON HEALTHCARE, INC. HEM-907XLINST Rev A...

Table of Contents