Endress+Hauser Liquistation CSF33 Brief Operating Instructions

Endress+Hauser Liquistation CSF33 Brief Operating Instructions

Automatic sampler for liquid media
Hide thumbs Also See for Liquistation CSF33:

Advertisement

Quick Links

KA01168C/07/EN/07.20
71473506
2020-03-19
Products
Brief Operating Instructions
Liquistation CSF33
Automatic sampler for liquid media
These instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Detailed information on the device can be found in the
Operating Instructions and in the other documentation
available at:
• www.endress.com/device-viewer
• Smart phone/tablet: Endress+Hauser Operations App
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Liquistation CSF33

  • Page 1 These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Detailed information on the device can be found in the Operating Instructions and in the other documentation available at: • www.endress.com/device-viewer • Smart phone/tablet: Endress+Hauser Operations App...
  • Page 2 Liquistation CSF33 Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0040778 Endress+Hauser...
  • Page 3: Table Of Contents

    Liquistation CSF33 Table of contents Table of contents About this document ............. . 4 Warnings .
  • Page 4: About This Document

    About this document Liquistation CSF33 About this document Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious injury. Causes (/consequences) If necessary, Consequences of non- compliance (if applicable) ‣...
  • Page 5: Symbols On The Device

    Reference to device documentation Documentation The following manuals which are available on the product pages on the internet complement these Operating Instructions: • Operating Instructions for Liquistation CSF33, BA00479 • Device description • Commissioning • Operation • Software description (excluding sensor menus; these are described in a separate manual - see below) •...
  • Page 6: Basic Safety Instructions

    Designated use Liquistation CSF33 is a stationary sampler for liquid media. The samples are taken discontinuously using a vacuum pump or peristaltic pump and are then distributed into sampling containers and refrigerated.
  • Page 7: Operational Safety

    Liquistation CSF33 Basic safety instructions Operational safety Before commissioning the entire measuring point: Verify that all connections are correct. Ensure that electrical cables and hose connections are undamaged. Do not operate damaged products, and protect them against unintentional operation. Label damaged products as defective.
  • Page 8: Incoming Acceptance And Product Identification

    Incoming acceptance and product identification Liquistation CSF33 Incoming acceptance and product identification Incoming acceptance Verify that the packaging is undamaged.  Notify the supplier of any damage to the packaging. Keep the damaged packaging until the issue has been resolved.
  • Page 9: Scope Of Delivery

    Liquistation CSF33 Incoming acceptance and product identification Scope of delivery The scope of delivery comprises: • 1 Liquistation CSF33 with: • The ordered bottle configuration • Optional hardware • Accessories kit • • Connection nipple for suction line with various angles (straight, 90°), Allen key (for version with vacuum pump only) •...
  • Page 10: Installation

    Installation Liquistation CSF33 Installation Installation conditions 4.1.1 Dimensions 625 (24.6) 471 (18.5) 753 (29.7) 355 (14.0) 125 (4.9) 816 (32.1) A0014539  1 Dimensions of Liquistation CSF33 plastic version, dimensions in mm (in) Suction line connection Endress+Hauser...
  • Page 11 Liquistation CSF33 Installation 624 (24.6) 467 (18.4) 743 (29.3) 355 (14.0) 125 (4.9) 816 (32.1) A0014550  2 Dimensions of Liquistation CSF33CSF33 stainless steel version, dimensions in mm (in) Suction line connection Endress+Hauser...
  • Page 12 Installation Liquistation CSF33 4.1.2 Installation site For version with pump A0024411  3 Liquistation mounting conditions Correct  The suction line must be routed with a downward slope to the sampling point. Incorrect  The sampler should never be mounted in a place where it is exposed to aggressive gases.
  • Page 13 Liquistation CSF33 Installation • Protect the device against strong magnetic fields. • Make sure air can circulate freely at the side panels of the cabinet. Do not mount the device directly against a wall. Allow at least 150 mm (5.9") from the wall to the left and right.
  • Page 14 Installation Liquistation CSF33 4.1.4 Connection for suctioning samples • Maximum suction height: • Vacuum pump: Standard 6 m (20 ft) • Peristaltic pump: standard 8 m (26 ft) • Maximum hose length: 30 m (98 ft) • Hose connection diameter •...
  • Page 15 Liquistation CSF33 Installation Requirements for the sampling point: • Do not connect the suction line to pressurized systems. • Use the suction filter to impede coarse and abrasive solids and solids which can cause clogging. • Immerse the suction line in the direction of flow.
  • Page 16: Installation

    Installation Liquistation CSF33 Installation 4.2.1 Connecting the suction line at the side on version with pump When installing the device, take the installation conditions into account. Route the suction line from the sampling point to the device. Screw the suction line onto the device' s hose connection.
  • Page 17 Liquistation CSF33 Installation Connecting the suction line on version with vacuum pump A0013708 A0013707  7 Suction line connected from below  6 Connecting the suction line from the side (as- delivered state) Hose Fixing clip for hose gland Thread adapter nut Hose gland Unscrew the thread adapter nut (item 3).
  • Page 18: Post-Installation Check

    Installation Liquistation CSF33 Connecting the suction line on version with peristaltic pump A0013705 A0013706  8 Connecting the suction line from the side (as-  9 Suction line delivered state) Small thread adapter nut Hose Thread adapter nut Hose gland Unscrew the thread adapter nut (item 3) and the hose fitting (item 4) from the side panel.
  • Page 19: Electrical Connection

    Liquistation CSF33 Electrical connection Electrical connection Connecting the sampler WARNING Device is live! Incorrect connection may result in injury or death! ‣ The electrical connection may be performed only by an electrical technician. ‣ The electrical technician must have read and understood these Operating Instructions and must follow the instructions contained therein.
  • Page 20 Electrical connection Liquistation CSF33 5.1.3 Removing the rear panel of the dosing compartment Open the door of the dosing compartment. Using an 5mm Allen key, release the rear panel by turning the lock clockwise.  CLOSE OPEN A0012803 Lift up the rear upper panel and pull it off towards the back.
  • Page 21 Liquistation CSF33 Electrical connection 5.1.4 Removing the rear panel of the sampling compartment A0012825 ‣ Remove the bolt on the rear of the dosing compartment.  A0012824 ‣ Remove the bolt on the rear panel. 5.1.5 Removing the cover WARNING...
  • Page 22 Electrical connection Liquistation CSF33 A0012831 Release the screw with an Allen key (5 mm). Remove the cover of the power unit from the front. When reassembling make sure that the seals are seated correctly. 5.1.6 Power supply terminal assignment The power supply is connected via plug-in terminals.
  • Page 23 Liquistation CSF33 Electrical connection A0013237  11 Terminal assignment Assignment: 100 to 120 V/200 to 240 V AC ±10 % Assignment: 24 V DC +15/-9 % Rechargeable batteries (optional) Internal 24 V voltage Fuses (only for batteries) 5.1.7 Cable terminals After connection, make sure that every cable end is securely in place.
  • Page 24 Electrical connection Liquistation CSF33  12 Press the screwdriver  13 Insert the cable until the  14 Remove the screwdriver against the clip (opens limit stop (closes the terminal) the terminal) Endress+Hauser...
  • Page 25: Connecting Modules And Sensors

    Liquistation CSF33 Electrical connection Connecting modules and sensors 5.2.1 Connection compartment in the controller housing A0042244 A0012843 1 E base module Sampler controller The controller housing has a separate connection Display cover open, version with base module E compartment. Release the six cover screws to open the connection compartment: ‣...
  • Page 26 Electrical connection Liquistation CSF33 5.2.2 Base module SYS BASE2-SYS Status A0042245  15 Base module SYS (BASE2-SYS) SD card slot Slot for display cable Ethernet interface Connecting cable to sampler controller Voltage connection Service interface Internal device connection, do not disconnect the plug.
  • Page 27 Liquistation CSF33 Electrical connection Analog inputs and binary inputs/outputs 125 123 124 225 223 224 191 192 291292 145 146 245 246 A0042282  16 Position of the terminals Analog inputs 1 and 2 Binary inputs/outputs Analog inputs A0012989  17...
  • Page 28 Electrical connection Liquistation CSF33 CSXxx CMxx A0028652  18 With two-wire transmitter, e.g. Liquiline M CM42 CSXxx FMU90 A0028653  19 With four-wire transmitter, e.g. Prosonic S FMU90 Binary inputs A0013381  20 Assignment of binary inputs 1 and 2...
  • Page 29: Terminal Assignment For Input/Output Signals

    Liquistation CSF33 Electrical connection When connecting to an internal voltage source, use the terminal connection on the rear of the dosing compartment. The connection is located on the lower terminal strip (on the far left, + and -), (→  22)
  • Page 30: Ensuring The Degree Of Protection

    Electrical connection Liquistation CSF33 Ensuring the degree of protection Only the mechanical and electrical connections which are described in these instructions and which are necessary for the required, designated use, may be carried out on the device delivered. ‣ Exercise care when carrying out the work.
  • Page 31: Post-Connection Check

    Liquistation CSF33 Electrical connection Post-connection check WARNING Connection errors The safety of people and of the measuring point is at risk! The manufacturer does not accept any responsibility for errors that result from failure to comply with the instructions in this manual.
  • Page 32: Operation Options

    Operation options Liquistation CSF33 Operation options Overview 6.1.1 Display and operating elements Display (with red display background in alarm condition) Navigator (jog/shuttle and press/hold EH_CSF 09:11:05 31.03.2015 function) Soft keys (function depends on menu) MODE A0025501  24 Overview of operation 6.1.2...
  • Page 33: Access To The Operating Menu Via The Local Display

    Liquistation CSF33 Operation options Access to the operating menu via the local display 6.2.1 Operating concept MODE MODE Turningthe navigator: moving the cursor in the menu Pressingthe soft key: selecting the menu directly Menu/Language MODE Turningthe navigator: selecting a value (e.g. from a list)
  • Page 34: Configuration Options

    Operation options Liquistation CSF33 6.2.2 Locking or unlocking operating keys Locking operating keys ‣ Press the navigator for longer than 2 s.  A context menu for locking the operating keys is displayed. You have the choice of locking the keys with or without password protection. "With password"...
  • Page 35 Liquistation CSF33 Operation options 6.3.3 Numerical values • You are changing a variable. • The maximum and minimum values for this variable are shown on the display. • Configure a value within these limits. • Example: Menu/Operation/Display/Contrast 6.3.4 Actions • You trigger an action with the appropriate function.
  • Page 36 Operation options Liquistation CSF33 6.3.5 Free text • You are assigning an individual designation. • Enter a text. You can use the characters in the editor for this purpose (upper-case and lower-case letters, numbers and special characters). • Using the soft keys, you can: •...
  • Page 37 Liquistation CSF33 Operation options 6.3.6 Tables • Tables are needed to map mathematical functions or to enter irregular interval samples. • You edit a table by navigating through rows and columns with the navigator and changing the values of the cells.
  • Page 38: Commissioning

    Commissioning Liquistation CSF33 Commissioning Function check WARNING Incorrect connection, incorrect supply voltage Safety risks for staff and device malfunctions! ‣ Check that all connections have been established correctly in accordance with the wiring diagram. ‣ Ensure that the supply voltage matches the voltage indicated on the nameplate.
  • Page 39 Liquistation CSF33 Commissioning 7.3.2 Display behavior Menu/Operation/Display Function Options Info Contrast 5 to 95 % Adjust the screen settings to suit your working environment. Factory setting 50 % Backlight = Automatic The backlighting is switched off automatically Backlight Selection after a short time if a button is not pressed. It •...
  • Page 40 Commissioning Liquistation CSF33 7.3.3 User definable screens Menu/Operation/User definable screens Function Options Info Meas. screen 1 ... 6 You can create 6 measuring screens of your own and give them a name. The functions are identical for all 6 measuring screens.
  • Page 41 Liquistation CSF33 Commissioning 7.3.4 Basic setup Making basic settings Switch to the Setup/Basic setup menu.  Make the following settings. Device tag: Give your device any name of your choice (max. 32 characters). Set date: Correct the set date if necessary.
  • Page 42 Commissioning Liquistation CSF33 Standard (1 sampling program with 1-5 sub-programs) Start condition: • Immediate activation, individual Stop condition: • Immediate times, multiple times, interval, • Program end • Date/time deactivation of sub-program 1 • Continuous • Volume • Time-paced, volume-paced or flow-...
  • Page 43 Liquistation CSF33 Commissioning Manual sampling Menu/Manual sampling 15000 ml Distribution position Bottle 1 Multiplier MODE A0036865-EN Manual sampling is triggered by the MAN soft key. This pauses any program currently running.  The current bottle configuration and the current sample volume are displayed. You can select the distributor position.
  • Page 44 Commissioning Liquistation CSF33 Select the Bottle change mode after number of samples or time for average samples. With the option "Bottle change after a time", you can enter the change time and bottle synchronization (None, 1st bottle change time, 1st time of change + bottle number). The description for this can be found in the "Bottle synchronization"...
  • Page 48 *71473506* 71473506 www.addresses.endress.com...