Attaching the swivel-arm covers

Attaching the swivel-arm covers

Arranging the swivel-arm system

1. Install object table, no compression force (swivel-arm system still upside down).
2. Remove the protective strips from the metal curtain.
3. Run the lifting carriage upwards.
4. Put back the protective strips onto the metal curtain.
5. Rotate the swivel-arm system to 0 degrees by pressing one of the switches for
6. System and mains "OFF".
7. Remove the swivel-arm system transport safeguard, see Chapter 3

Connecting the cables to control-button boards and patient handles

MAMMOMAT 1000/3000 Nova
Fig. 1 Connecting the cables
Before installing the side covers, the cables are to be connected.
1. Connect cables X807 (1) to control-button circuit boards D807 (2).
2. Connect ground wires (3) to patient handles (4).
Siemens AG
Medical Solutions
clockwise rotation on the supporting arm.
"Preparatory work".