Cable connections
EMC measures on the bottom plate
Fitting the cables X1, X8 and X9 onto the bottom plate
3
Fig. 3 Bottom plate
1. Fit the bottom plate at the bottom of the generator stand as shown in Fig. 3.
2. Carefully cut away the insulation on cables X1, X8 and X9 to lay bare the shield
3. Select proper clamps (2) with regard to the thickness of the cables. Cable clamps
4. Fit the clamp (2) over the shield braid and screw tight to ensure satisfactory
NOTE
5. Fasten the bottom plate with four screws (3) and contact washers (4).
NOTE
NOTE
Siemens AG
Medical Solutions
2
4
braid (1).
in different sizes are provided on the bottom plate.
contact between shield braid and clamp.
Make sure that all excess cable lengths are placed below the bot-
tom plate.
Cable X9 shall also be clamped to the side wall of the cabinet, if
not already done, see Fig. 7 in this chapter.
Do not remove insulation and do not ground the shields of the
AEC signal cables to X10 and X11.
Register 3
1
SPB7-230.033.11
Rev. 06
05.06
Page 3 of 14
M1000/3000 Nova
CS PS SP
Installation and Start-Up