Chapter 4
SITE PREPARATION BEFORE INSTALLATION
IMPORTANT
Prior to placing the order and before installation, carefully check the
following requirements for the x-ray room.
Standard Compliance
Install the unit in an x-ray room compliant with all official regulations applicable to
protection against radiation. This room must reduce the frequency interferences of the
30MHz to 1GHz band to at least 12db.
Environmental Requirements
Check the following ambient operating condition requirements of the x-ray room before
installing the unit:
•
Temperatures: 5 ~35 °C
•
Relative humidity: 30 ~ 85%
•
Atmospheric pressure: 700 ~ 1060 hpa
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02