TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on
Batteries have run down.
when the START/STOP
button is pushed.
Battery polarities have
been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on display
incorrectly.
or the blood pressure
value is displayed
Did you talk or move
excessively low (or high).
during measurement?
Shaking of the arm with
the cuff on.
Er error symbol shown on
Did you move during
display - measurement
measurement?
result exceeds
Shaking of the arm with
measurement range.
the cuff on.
E1 error symbol shown
Air circuit abnormality.
on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on display
exceeding 300 mmHg.
E3 error symbol shown
Data error.
on display
CA appears on display
Recalibration alert.
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on the
warranty page.
CORRECTION
Replace them with two
new AA alkaline batteries.
Re-insert the batteries in
the correct positions.
Wrap the cuff properly so
that it is positioned
correctly.
Measure again. Keep arm
steady during
measurement.
Measure again. Do
not move during
measurement.
Check cuff connection.
Measure again.
Turn the unit off, then
measure again.
Remove and reload the
batteries.
Refer to page 25 for more
information.
29

SPECIFICATIONS

Power Source:
Two 1.5V DC (AA) batteries
Measurement
Oscillometric
Method:
Pressure: 40~250 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Manual Inflation
Deflation:
Automatic Pressure release valve
Memory Capacity:
60 memories
Auto-shut-off:
1 minute after last button operation
Operation
Temperature: 50°F~104°F (10°C ~40°C)
Environment:
Humidity: 40~85% RH max.
Storage
Temperature: 14°F~140°F (-10°C~60°C)
Environment:
Humidity: 10~90% RH max.
Weight:
.30 lbs/338 grams (without batteries)
Standard size adult arm cuff:
Arm Circumference:
9.5"–14.2" (24 – 36 cm)
4.50"(L) x 4.17"(W) x 2.25"(H)
Dimensions:
112mm(L) x 106mm(W) x 58mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
Note: These specifications are subject to change without notice.