Konica Minolta DiMAGE Scan Elite 5400 II Technical/Repair Manual

Hide thumbs Also See for DiMAGE Scan Elite 5400 II:

Advertisement

Quick Links

SERVICE MANUAL
DiMAGE Scan Elite 5400 II
DiMAGE Scan Elite 5400 II
(2892-100)
COMMON MODEL :
TECHNICAL ILLUSTRATION
PARTS NOTICE WITH PARTS LIST
TECHNICAL REPAIR MANUAL
KONICA MINOLTA PHOTO IMAGING, INC.

Advertisement

Table of Contents
loading

Summary of Contents for Konica Minolta DiMAGE Scan Elite 5400 II

  • Page 1 SERVICE MANUAL DiMAGE Scan Elite 5400 II DiMAGE Scan Elite 5400 II (2892-100) COMMON MODEL : TECHNICAL ILLUSTRATION PARTS NOTICE WITH PARTS LIST TECHNICAL REPAIR MANUAL KONICA MINOLTA PHOTO IMAGING, INC.
  • Page 2 (2892) 1 Specifications Type Scanner type 35mm film scanner Type 3-line color CCD Image sensor Total pixels 5340 pixels/ line Filter Primary color filter Scanning Film type 35mm film (Color/ B&W, Negative/ Positive) system Scan type Moving film, fixed sensor, single-pass scan 5400x5400 dpi Optical resolution Maximum input pixels...
  • Page 3 (2892) 1 INDEX INDEX INDEX INDEX ...... 2892-0110 ...... 2892-0111 ...... 2892-0113 ..... 2892-0401 ..... 2892-0402 2892-0116 ....... 2892-1002 ..... 2892-1004 ..... 2892-1005 ..... 2890-1006 ..... 2890-1007 ..... 2890-1008 ..... 2890-1009 ..... 2890-1201 ..... 2890-1303 2892-0230 .
  • Page 4 1(2892) 2892-1004-01 2892-1002-01 2886-9626-01 (2890-1008-01) (2890-1007-01) 2890-4005-01 2892-0401-01 2892-0110-01 2890-4003-01 9384-2191-00 (4) 2890-9003-01 (3) (10x20mm) 2892-0230-01 2890-9002-01 (2) (2890-1009-01) (2890-1201-01) (2890-1006-01) 2892-0116-01 (2892-4006-01) 2890-0091-01 2890-0065-01 2890-9005-01 (2) (2890-4002-01) 2892-0402-01 2892-0113-01 2890-9002-01 (2) 2892-0111-01 2890-1303-01 (2) 2890-9004-01 (4) 2890-1303-01 (2) 2886-9823-01 (2) 2892-1005-01...
  • Page 5 (2892) 1 PART NO PART NAME QTY. 撮像ボディユニット 2892-0110-01 IMAGE TAKING BLOCK フォーカスダイアル (2890-1006-01) FOCUS DIAL アーム (2890-1007-01) ARM 電源SW SP (2890-1008-01) POWER SW. SP ギア12T (2890-1201-01) GEAR 12T FFCケーブル8P (2892-4006-01) FFC CABEL 8P 前カバーセット 2892-0111-01 FRONT HOUSING ASSY 電源ボタン (2890-1009-01) POWER BUTTON ケーブル6P...
  • Page 6 COMMON REASON MODEL →○(×) (○)×← Previous (10) (11) (12) Added./新設 1234-5678-90 (Parts No.) ABC ASSY (Part name) ABC セット(部品名称) (13) セットでの互換性/ Interchangeability as a set. Previous →○(×) (○)×← (14) (15) Continued on the next page./次ページにつづく KONICA MINOLTA PHOTO IMAGING, INC.
  • Page 7 N E W / 新規設定 新規設定 新規設定 N E W / N E W / N E W / N E W / ■本案内状に従い、パーツリストへの転記等、メンテナンスを行なってください。 According to this list, correct Parts List in Service Manual for effective parts management. KONICA MINOLTA PHOTO IMAGING, INC.
  • Page 8 新規設定 新規設定 新規設定 新規設定 N E W / N E W / N E W / N E W / ■本案内状に従い、パーツリストへの転記等、メンテナンスを行なってください。 According to this list, correct Parts List in Service Manual for effective parts management. KONICA MINOLTA PHOTO IMAGING, INC.
  • Page 9 (2892) REPAIR GUIDE Contents of this manual are in accordance with the assembly procedure. Therefore, follow the reverse procedure when disassembling. SYMBOLS : Caution and key points of assembly : Anti-diffusion agent : Adhesive : Grease : Tools TABLE OF CONTENTS Page Image taking block/ Main PCB Assy/ Exteior Parts Installation ................
  • Page 10 (2892) Image taking block/ Main PCB Assy/ Exterior Parts Installation 1 0 0 4 1 0 0 4 1 0 0 4 1 0 0 4 1 0 0 4 UPPER CASE UPPER CASE UPPER CASE UPPER CASE UPPER CASE 2 8 8 6 - 9 6 2 6 - 0 1 2 8 8 6 - 9 6 2 6 - 0 1 2 8 8 6 - 9 6 2 6 - 0 1...
  • Page 11 (2892) ■ ■ ■ ■ ■ Fig. 1 Fig. 1 Fig. 1 Fig. 1 Fig. 1 Put through a hole of Main PCB. Put through a hole of Main PCB. Put through a hole of Main PCB. Put through a hole of Main PCB. Put through a hole of Main PCB.
  • Page 12 (2892) ■ ■ ■ ■ ■ Fig. 4 Fig. 4 Fig. 4 Fig. 4 Fig. 4 Connect the connector by the position as shown. Connect the connector by the position as shown. Connect the connector by the position as shown. Connect the connector by the position as shown.
  • Page 13 (2892) ■ ■ ■ ■ ■ Fig. 5 Fig. 5 Fig. 5 Fig. 5 Fig. 5 Focus dial Assembly Procedure. DiMAGE Scan Utility is necessary for this procedure. Please install "DiMAGE Scan" Software from CD-ROM to your PC. Preparations 1. Remove Upper case, Front housing assy, focus dial and gear 12T. Fig.
  • Page 14 (2892) Measuring Instruments, Tools and Jigs Slide Mount Holder (SH-M20) DiMAGE Scan Elite 5400 II (CD-ROM) AC Adapter PC/AT compatibles CPU: Pentium III or later RAM: 500MB or more HDD: Approx. 2GB or more of available hard-disk space OS: Windows 98, 98SE, Me, 2000 Professional, XP Professional, XPHE Monitor: 1024 x 768 (pixels) is recommended.
  • Page 15 1 Approved by Verified by Written by Division Camera CS Division Issued July 15, 2005 Yata Ueno Approved Verified Written Issued/Code/Version/Page Title Repair information for distortion of ’05.07.15 QS FA 2892-S01E Yata Ueno index images KONICA MINOLTA PHOTO IMAGING, INC.
  • Page 16: Contents

    CODE NO. 2892-100 SERVICE MANUAL SUPPLEMENTARY MODEL DiMAGE Scan Elite 5400 II INFORMATION Repair information for distortion of index images Contents With 35mm film holder, the subject's slanting line in the 6th (or 4th) frame of index image may be distorted like wave.
  • Page 17: Table Of Contents

    (2892) 1 CHECK LIST 1. This Check List describes the quality of operation warranted to general users. When users inquire about quality or request inspection, refer to this Check List. Use this list also when checking operation after repair. 2. When using this list at shipping or receiving inspection, judge the quality according to the purpose of the inspection, not by directly referring to this level.
  • Page 18: Performance

    Check the scanner with the computer which meets the system requirements in the last page of this check list. The driver software DiMAGE Scan Elite 5400 II supplied with the product is necessary for check. If this driver is not installed in the computer, please install it beforehand.
  • Page 19: Setting The Chart

    (2892) 3 Setting the chart Make sure the chart faces the CCD side in the Slide Mount Holder (SH-M10) as illustrated. Up: Fluorescent Front view Side view Down: CCD Scan 1) Change the setting of "Glass Chart" of "DiMAGEScan.ini" located in the followong folder to "1"and save it (C:\Program Files\DS_Elite5400_2) 2) Click Start >...
  • Page 20: Measurement

    4 (2892) Measurement 1) Select Start > Program > DPEV > GEOM. 2) Click Setup > Measure and set the preference as below. Measuring condition Size Auto Detect Resolution 5400 dpi Chart Size 20/30/40 Measure item Check all Data Out Auto Save Pasted MTF-Data Click the Profile “New”...
  • Page 21: Check

    (2892) 5 Check 1) Open 2892 data sheet.xls in the FD and select the worksheet “Scanner Gauge”. 2) Move the mouse pointer to the cell F3, and click Edit > Paste. The file name is inputted in F3 and the performance data is inputted in the column F. 3) Confirm that the cell G3 is null.
  • Page 22: Color Chart

    Microsoft Excel Setup 1. Confirm the driver software DiMAGE Scan Elite 5400 II is launched or not. If not installed yet, please install the driver software first. Be sure to remove the DiMAGEScan.ini from the folder after scanning the chart.
  • Page 23: Scan

    (2892) 7 Scan 1) Click Start > Programs > DiMAGE Scan Elite5400 2 Ver. 1.* > DiMAGE Scan Elite5400 2 Utility. 2) Set the preferences as below. Scanning condition Film Type Color Positive Input resolution 540 dpi 3) Click on the Prescan button to prescan the chart. Check that the chart is scanned as shown and that it is not underexposure.
  • Page 24: Standard

    8 (2892) Measurement condition Category Scanner Resolution 540 dpi Chart Standard 90Color135.txt Data Out Auto Save File Group Locate Vert:0 Hriz:0 Scale: 1 Sample Area Gamma Light Source D65 (Std) 3) Click “New” of the Profile and input a profile name, and then click on the Save button. The setting will be effective from the next measurement.
  • Page 25: Check

    (2892) 9 Check 1) Open 2892data sheet.xls in the FD and select the worksheet “Color Chart”. 2) Move the mouse pointer to the cell F3, and click Edit > Paste. The file name is inputted in F3 and the performance data is inputted in the column F. 3) Confirm that the cell G3 is null.
  • Page 26: Color Chart

    10 (2892) Frequency Characteristic Range Vertical 50 % (40 lines/mm) or greater Center Horizontal 50 % (40 lines/mm) or greater Vertical 40 % (40 lines/mm) or greater Around Horizontal 40 % (40 lines/mm) or greater Color chart Output Density Chart Range White 95 ±...
  • Page 27: Equipment Required

    2892Adjustment Program CD Version 1.0 <2892-0001-75> DiMAGEScan.ini 90Color135.dll 2892_data sheet.xls 2887Adjustment Program CD Version 2.0 <2887-0001-76> Scanner Gauge Chart (35mm) <7982-5002-01> Color Chart Slide (35mm) <7982-5005-01> Slide Mount Holder (SH-M20) Microsoft Excel DiMAGE Scan Elite 5400 II (CD-ROM) AC Adapter...

Table of Contents