3.2 Elimination of Test Leads Impedance Error

Relative operation is recommended in low resistance measurement, which can eliminate the test lead
impedance error.
Steps:
1)
Press
button to choose 2-wire resistance measurement.
2)
Connect the ends of test lead with the multimeter. Connect the red test lead to terminal Input-HI and
the black one to terminal Input-LO.
3)
Select proper range of resistance according to measuring resistance. Auto range is selected by
default.
4)
Short circuit the test leads and the resistance impedance will be displayed as Figure 3-3.
→ [Relatively] to set the reference value by direction buttons.
5)
Press
6)
Users can also open the relative value in measurement to obtain the test lead impedance after
relative operation.
Figure 3-3 Reading of short-circuit test leads

3.3 DBm Measurement

dBm measurement is commonly used in audio signal measurement.
Steps:
1)
Press
to enable AC voltage measurement and select proper voltage range.
7)
Connect a test lead with the multimeter. Connect the red test lead to terminal Input-HI and the black
one to terminal Input-LO. See Figure 3-5.
2)
Set parameters of dBm operation. Press
operation set value 50Ω as the reference value of hypothetical circuit. The power value of reference
resistance will be displayed.
Figure 3-4 Reading after relative operation
→ [dBm] to enable dBm function. Set the dBm
Figure 3-5 dBm measurement
45
UT8805E Benchtop Digital Multimeter