SECTION 5—REMOVING THE CONSERVING REGULATOR
SECTION 5—REMOVING THE
CONSERVING REGULATOR
NEVER attempt to remove the oxygen conserving regulator 
from an oxygen cylinder unless the cylinder valve is 
completely closed.
DO NOT try to loosen the T‐handle beyond the stop. 
Otherwise, damage to the oxygen conserving regulator will 
occur.
NOTE: For information on how to operate the cylinder, consult your 
equipment provider.
1. Completely close the oxygen cylinder valve.
2. Turn the oxygen conserving regulator selector switch to 
2 L/min. continuous flow position.
3. Wait for oxygen to stop flowing from the oxygen 
conserving regulator.
4. Turn the T‐handle to loosen the connection to the post 
valve.
5. Remove the oxygen conserving regulator from the post 
valve.