SECTION 3—INSTALLING THE CONSERVING REGULATOR
NOTE: For information on how to operate the cylinder, consult your
equipment provider.
NOTE: For this procedure, refer to FIGURE 3.1.
1. Loosen the cylinder post valve T‐Handle.
2. Lower the conserver over
any CGA870 post valve
cylinder with the
alignment pins toward the
holes on the cylinder neck.
3. Align the two pins and
gasket with the
corresponding holes on
the cylinder post valve.
NOTE: Cylinder post valve T‐Handle should be aligned with
indentation on post valve.
4. While holding the unit in place, tighten the cylinder post
valve by turning clockwise.
5. Attach a standard cannula [7 ft. (2.13 m) or less in length] to
the oxygen supply outlet fitting.
NOTE: Tighten only by hand. The use of a tool to tighten the T‐Handle
may damage the unit.
Part No 1160840
FIGURE 3.1 Attaching the
IOC100P to the Cylinder
13