SECTION 2—FEATURES
Selector Switch: Enables the user to select the desired setting, as 
well as "OFF" and "CF" (continuous flow). When not in use, the 
switch should be turned to the "OFF" position.
Oxygen Supply Outlet: Use this fitting to attach a standard 
cannula.
Cylinder Alignment Pins: When assembling the unit, these 
parallel pins MUST be inserted into the holes on the CGA870 
post valve.
CGA870 Seal Washer [Metal Washer with Viton
 or equivalent 
®
Seal]: Creates the interface between the post valve and the 
conserver regulator. Besides offering a rugged interface, it also 
surrounds the oxygen path in a ring of stainless steel or brass.
Vent Hole: Maintains proper internal pressure. DO NOT 
obstruct with any object, such as a label or tight‐fitting carrying 
bag.
NOTE: Use only a CGA870 Seal Washer [Metal Washer with Viton 
Seal]. DO NOT use a plastic compression gasket.
FIGURE 2.2 Connection view of the IOC100P with CGA870
Connection
Part No 1160840
11