SECTION 6—TYPICAL PRODUCT PARAMETERS
SHOCK VIBRATION:

OXYGEN CYLINDER DURATION

The total delivery of oxygen via the conserver is related to the 
breathing rates, it is user adaptive in that the total oxygen 
delivered per minute will automatically adjust with user need, as 
expressed by increased or decreased breathing rates. For 
example, at all settings, twice as much oxygen per minute will be 
delivered if one breathes twenty (20) times per minute as 
compared with ten (10) times per minute. Refer to OXYGEN 
DURATION TABLE below.
OXYGEN DURATION TABLE
SETTING
Cylinder Type
Cylinder Volume
Liters @ 2015 psi
M2
M4(A)
M6(B)
ML6
M9(C)
D
E
Invacare® Element™
1
2
36
1.9
113
5.9
2.9
164
8.5
4.3
171
8.9
4.5
246
12.8
6.4
425
22.1 11.1
680
35.4 17.7 11.8
Not to exceed IEC 601-1 requirements
Not to exceed IEC 68-2-6, IEC 68-2-34
No smoking or open flames
Consult accompanying documents
3
4
5
Estimated Cylinder Duration in Hours
(Based on 20 breaths/min.)
.9
.6
.5
.4
2.0
1.5
1.2
2.8
2.1
1.7
3.0
2.2
1.8
4.3
3.2
2.6
7.4
5.5
4.4
8.9
7.1
20
6
Continuous
Flow
2 L/min.
.3
.3
1.0
.9
1.4
1.4
1.5
1.4
2.1
2.1
3.7
3.5
5.9
5.7
Part No 1160840