Chapter XIV Technical Information

Unless otherwise specified, all technical specifications are applicable to the probes
with attenuation switch set as 10× as well as UTD2000 series DSO. DSO must first meet
the following two conditions to meet those specification standards:
The instrument must continuously operate for over half an hour with in the operating
temperature.
If the operating temperature has exceeds 5℃, please perform the
"Self-Adjustment "function in the UTILITY system function.
Any of the specification with "Typical" value represented is for reference only。
Technical Specification
Acquire System Specification
Model
UTD2025CL
Real-time
250MS/s
Equivalence
——
N can be chosen among 2, 4, 8, 16, 32, 64, 128 and 256; N is the number of average
Average
samples.
Input Channel Specifications
Input Coupling
DC, AC and GND
Input impedance
(1MΩ± 1%)//(24pF± 3 pF)
Probe attenuation
1×, 10×, 100×, 1000×
coefficient
Maximum input
400V (DC+AC peak value, 1MΩ input impedance)
voltage
Horizontal System Specification
Scanning scope
Waveform interpolation
ampling rate and delay time
accuracy
Record length
Storage depth
Sampling rate and delay
time accuracy
Measurement accuracy of
time interval (△T) (full
bandwidth)
Vertical
UTD2052CL
UTD2102e
500MS/s
500MS/s
25GS/s
25GS/s
5ns/div-50s/div
Sin(x)/x
±50ppm (any time interval≥1m)
2×512k sampling point
Single channel: 25k;Double channel: 12.5k
±50ppm(any time interval ≥1ms )
Single time: ± (1 sampling time interval+50ppm×reading+0.6ns)
>16 average values: ± (sampling time interval+50ppm×reading+0.4ns)
56
UTD2000 Series User Manual
UTD2052CEX
UTD2072CEX
1GS/s
50GS/s
UTD2102CEX