Omega Engineering i Series User Manual page 21

Process/strain gauge monitor process/strain gauge limit alarm
Table of Contents

Advertisement

ENTER INPUT TYPE MENU:
Press a
Press d
Press d
INPUT TYPE MENU:
Press b
Press d
Input Types:
Display:
To have +100 mV you need to connect to 0-1 V.
RATIOMETRIC OPERATION SUBMENU:
Press d
Press b
Press d
The Ratiometric operations are typically used for Strain gauge monitor.
If your instrument is configured as Process (voltage and current), set RT1B
to DSBL disable Ratiometric operations.
If ENBL Ratiometric operations Enabled was selected, the changes to
the excitation voltage will be compensated through Ratio measurement.
If DSBL Ratiometric operation Disabled was selected, any changes to
the excitation voltage will effect the output of strain gauge bridge and, as
a result, a reading of the instrument.
DISPLAY RESOLUTION SUBMENU:
Press a
Press d
Press d
If LO Low Resolution was selected the resolution of the display is 10 V.
If HI High Resolution was selected the resolution of the display is 1 V. In
case of High Resolution, the maximum input signal is 10 mV.
1) Press a, if necessary, until CNFG prompt appears.
2) Display advances to INPT Input Menu.
3) Display flashes 0-0.1 , 0-1.0 , 0-10 or 0-20 (0 to 100 mV,
0 to 1 V, 0 to 10 V or 0 to 20 mA).
4) Scroll through the available selection of input ranges 0-0.1 ,
0-1.0 , 0-10 or 0-20 to the selection of your choice.
5) Display shows STRD stored message momentarily and then
advances to the RT1B Ratiometric operation submenu.
100 mV 1 V
0-0.1
0-1.0
6) Display flashes previous selection of ENBL Enable or DSBL
Disable.
7) Scroll through the available selection ENBL or DSBL
(flashing).
8) Display shows STRD stored message momentarily and then
advances to RESO Display Resolution Submenu.
9) Display flashes previous selection of LO Low or HI High
resolution.
10) Scroll through the available selection LO or HI (flashing).
11) Display shows STRD stored message momentarily and then
advances to BUTN Button Selection Submenu.
10 V
0 – 20 mA
0-10
0-20
17

Advertisement

Table of Contents
loading

Table of Contents