DIAGNOSTICS
10DEC05
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
DG4825-1
Page
Action:
98 of 180
Cause 2: The PMT/DAS BOARD A5 malfunctioned.
Action:
16013 - Imaging System contains out-of-range noise
Description: The "peak-to-peak" signal was high when the CR 825/850 SYSTEM was
Cause 1: The PMT/DAS BOARD A5 malfunctioned.
Action:
Cause 2: Noise in the POWER SUPPLY PS1.
Action:
16014 - Imaging System did not zero or PMT did not calibrate
Cause 1: The DOSIMETER malfunctioned.
Action:
Cause 2: The exposure on the PLATE is not correct during the calibration.
Action:
Cause 3: The X-RAY TUBE malfunctioned.
Action:
Do the checkout for the
Do the checkout for the
set to "AUTO-ZERO."
Do the checkout for the
Do the checkout for the
Check that the DOSIMETER operates correctly.
Do the calibration with another PLATE.
Expose the PLATE using a new X-RAY TUBE.
MASTER CENTRAL PROCESSING UNIT BOARD
PMT/DAS BOARD
A5.
PMT/DAS BOARD
A5.
POWER SUPPLY
PS1.
Error Codes
A2.