Agilent Technologies VS C15 Series Field Replacement Instructions page 7

Calibrated leak replacement for leak detectors

Advertisement

6. Remove the assembly and use a flat head screw driver to remove the two screws holding the calibrated leak
(Figure 4: Calibrated Leak Screws).
7. Discard the calibrated leak and the O-ring.
8. Inspect the replacement calibrated leak and O-ring for damage or particle contamination and remove any
particle contamination.
9. Install the O-ring in the groove and then assemble the replacement calibrated leak to the valve block using a
slotted screwdriver.
10. Connect the calibrated leak temperature cable end labeled P1A to the leak assembly board connector J1A
(Figure 5: Leak Assembly Board Connection).
Calibrated Leak Replacement for VS C15, VS, and (G8610, G8611, or G8612) Series Leak Detectors
Figure 4: Calibrated Leak Screws
Figure 5: Leak Assembly Board Connection
Page 5 of 12