6.5
www.eltra-trade.com

Changing basic settings

In the basic settings menu it is possible to set the following parameters:
Parameter
Description
Main frequency
Selection of main power supply frequency corresponding to the country in
which the flowmeter is installed (e.g. 60 Hz in America).
Flow direction
Selection of correct flow direction in pipe.
Customer units
Setting of user defined volume and time units.
Q
Setting of measuring range, analog outputs and frequency output. Also
max
individual dimension-dependent setting of value, decimal point, unit and time.
Q
2
Setting of measuring range, analog outputs and frequency output. Also
max
individual dimension-dependent setting of value, decimal point, unit and time.
This menu is only visible if chosen as external digital input.
Totalizer
Setting of unit and decimal point.
Low flow cut-off
Setting of a percentage of selected Q
reducing fluctuations in display and all outputs.
Empty pipe cut-off
When set to "On" the alarm will indicate when sensor is running empty. All
readings, display and outputs, will indicate zero.
Velocity unit
Setting of velocity unit per time unit
Error level
Selecting error level at which flowmeter will detect an error.
Note
Totalizer 2 is not visible when batch is selected as digital output.
Note
Q
2 is visible only when chosen as digital input.
max
Change basic settings as follows:
1. Press top up key
2. Enter password.
3. Use forward key
4. Press lock/unlock key
5. Use forward key
6. Press lock/unlock key
7. Use select key
and change key
8. Press lock/unlock key
for 2 sec.
to reach basic settings menu.
to unlock settings.
or backward key
to reach relevant menu.
to unlock settings.
to change settings.
to confirm new settings.