Cable Requirements - Siemens SITRANS F M MAG 5000 Manual

Hide thumbs Also See for SITRANS F M MAG 5000:
Table of Contents

Advertisement

Technical data

11.5 Cable requirements

11.5
Cable requirements
Basic data
Max. cable loop resistance
Cable glands on sensor
72
www.eltra-trade.com
No. of conductors
Min. sqr. area
Screen
Max. capacitance
Media temperature:
< 100 °C
> 200 °C
M20x1.5 gland - Cable ø 5 ... 13 mm (0.20 ... 0.51 inches)
½ NPT gland - cable ø 5 ... 9 mm (0.20 ... 0.35 inches)
WARNING
Cable glands
For Ex zone 1 installations only certified cable glands with protection type "e" can be used
for the power supply and the coil cable. The cable glands must be approved for the actual
temperature and cable dimension.
0.5 mm
info@eltra-trade.com
Coil
Electrode
cable
cable
2
3
0.2 mm
2
Yes
Yes
N/A
350 pF/m
40 Ω
N/A
6 Ω
N/A
+421 552 601 099
2