Technical data
11.5 Cable requirements
11.5
Cable requirements
Basic data
Max. cable loop resistance
Cable glands on sensor
72
www.eltra-trade.com
No. of conductors
Min. sqr. area
Screen
Max. capacitance
Media temperature:
< 100 °C
> 200 °C
M20x1.5 gland - Cable ø 5 ... 13 mm (0.20 ... 0.51 inches)
½ NPT gland - cable ø 5 ... 9 mm (0.20 ... 0.35 inches)
WARNING
Cable glands
For Ex zone 1 installations only certified cable glands with protection type "e" can be used
for the power supply and the coil cable. The cable glands must be approved for the actual
temperature and cable dimension.
0.5 mm
info@eltra-trade.com
Coil
Electrode
cable
cable
2
3
0.2 mm
2
Yes
Yes
N/A
350 pF/m
40 Ω
N/A
6 Ω
N/A
+421 552 601 099
2