8. Export the file DB_ROM_RCP_MasterConfig.csv.
9. Open the .csv file in Notepad.
CAUTION! Do not open the file in other editors or Microsoft Excel.
10.If required, enter a new PLC IP address and PLC master ELAN address as shown in the
following example.
PLC IP address
PLC master ELAN address
Figure 3-4
Example for a DB_ROM_RCP_MasterConfig.csv file
11.Import the file.
12.Activate the changed configuration with the PLC web client or SIMATIC HMI.
13.Change the SIMATIC HMI network settings.
Changing the SIMATIC HMI network settings (Page 15)
14.Activate the PLC IP address at the front panel of the PLC by setting the CPU to "STOP" and then
back to "RUN".
15.Couple the PLC and SIMATIC HMI again.
16.Start the SITRANS AID IQ web server with updated IP address.
Calling the SITRANS AID IQ web server (Page 25)
3.4

Changing the SIMATIC HMI network settings

Requirement
You have connected all devices.
Analyzer Intelligence Director
Operating Instructions, 05/2022, A5E51328515-AA
3.4 Changing the SIMATIC HMI network settings
Commissioning
15